accessories and options
file: n:\article\cos14310m5-v1_08hpc-en\20150130_cos14310m5-v1.08hpc-en_instructions_for_use_h-p-cosmos_treadmill.doc
© 2015 h/p/cosmos sports & medical gmbh author: fh [email protected] created 30.01.2015 printed 30.01.2015 page: 190 of 216
Connect the printer (with PCL printer
language and with a serial interface) via
interface cable with the serial interface RS232
of the running machine.
Attention: For a printer with a parallel interface
you will need a commercial interface
converter, available in every computer store,
at your local dealer or from the manufacturer
(h/p/cosmos order no.
[cos10056]
).
If option 20: 3 is chosen (printer protocol see
list of options), data will be printed out as
shown in the left example.
12.21
POLAR heart rate measurement system
N
In the case of interference do not rely on the indicated values of the wireless heart rate
transmission.
N
Precision of heart rate measuring: + / - 1 %, resp. + / - 1 beat per minute.
N
Pay attention to the advice regarding safety and warnings about possible interferences in the
chapters on errors, technical data, EMC, and installation in this manual.
N
“WARNING! Heart rate monitoring systems may be inaccurate.
N
Over exercising, incorrect or forbidden use may result in serious injury or death.
N
If the user feels faint, sick, dizzy or pain or in case of any other health problems stop exercising
immediately and see a doctor.
You can use all POLAR transmitters (coded and uncoded) with your running machine. For running machines of
generation MCU 4.0 with firmware version lower than V 3.01.1, an EPROM update is available for the use of coded
transmitters. Note: Even when using the coded senders, the transmission of the heart rate to the treadmill is still un-
coded. The models h/p/cosmos locomotion, h/p/cosmos pulsar, h/p/cosmos venus and h/p/cosmos saturn are equipped
with the POLAR W.I.N.D. transmitter with WearLink technology and coded transmission. The transmitter is only
activated if placed correctly on the body. For correct placement of transmitter belt, see chapter 5 entitled “Operation”
and chapter 5.10.4 entitled "Cardio mode".