emergency stop
switch (safety
lanyard with
actuator & pull cord)
N
subject struggles, cord pulls out the
magnet
N
running belt & inclination system stop fast
N
machine is
NOT
cut off from main power
of drive motor and inclination system
N
display active, shows “PULL STOP”
N
interface active, but no start, speed or
elevation signal executed
N
keyboard inactive
The emergency stop switch (safety lanyard
with actuator, pull cord and clip) of the
150/50 LC (pluto) cuts off the main power -
no message is shown at the dispays
NOT
interrupted
during
“PULL STOP”
stage
N
reconnect magnet to
its position at
UserTerminal
N
press START key or
new command via
interface
Additional stop
button
for arm support
N
Stop button is pushed
N
running belt & inclination system stop fast
N
machine is
NOT
cut off from main power
of drive motor and inclination system
N
display active, shows “PULL STOP”
N
interface active, but no start, speed or
elevation signal executed
N
keyboard inactive
NOT
interrupted
during
“PULL STOP”
stage
N
release push force
from the button
N
press START key or
new command
through interface
Quick stop
by light barriers
(venus / saturn)
N
light beam is interrupted
N
running belt & inclination system stop
extremely fast
N
drive motor is used to stop the running belt
very fast
N
machine is
NOT
cut off from main power
of drive motor and inclination system
N
display active and shows “PULL STOP”
N
interface active, but no start, speed or
elevation signal executed
N
keyboard inactive
NOT
interrupted
during Quick
Stop stage
N
remove obstacle from
light beam
N
„pull-stop“ advice
stays for 10 seconds
N
press START key
or new command
through interface