OP90
Load or save default settings
With this option all option values (exceptions see below) can be
saved or reloaded.
With
a
or
d
the loading or saving is selected. After confirmation with
c
the safety request has to be confirmed again with
c
.
§ $ %
SAVE dEF
AULt
ArE
YOU
SUrE
SAV
Ed
LOAD dEF
AULt
ArE
YOU
SUrE
LOAd
Ed
If default values are supposed to be loaded without having been
saved before following message will appear and no change can be
performed.
nO d
EF S
AVEd
The key SW1 on the MCU 5 assembly board performs the same
action as the loading of default values with the following exception: if
there are no default values saved, the default values of the firmware
will be taken over.
Following option values are not handled as default values and remain
unchanged:
User - OP 02: Total distance
User - OP 03: Total time of operation
User - OP 04: Total time of use
Administrator-OP 10: Treadmill serial number
Administrator-OP 91: MCU total time of operation
Administrator-OP 92: Total distance
Administrator-OP 93: Total time of operation
Administrator-OP 94: Total time of use
Administrator-OP 95: MCU serial number
OP91
Total time of operation MCU (h)
= total time of switch on MCU
§
and
$
report: hours of operation
%
reports: h
This value can not be changed!
OP92* Total distance (km)
§
and
$
report: total distance in km
%
reports km
Difference to user option 02: the displayed value can also be
changed.