GETTING STARTED
27
6
CH2 EXT CC
CONTROL IN
Constant-current control input pin for
channel 2.
In the external voltage control mode, an
external voltage of 0V to 10V can be
used to control the ouput current from
~0A to the rated output current.
In the external resistance control mode,
an external reistance of 0kΩ to 10kΩ can
be used to control the output current
from ~0A to the rated output current.
7
CH2 COM
COM pin for channel 2.
Connected to the negative output
terminal of channel 2.
8
CH2 EXT V MON
OUT
Output voltage monitor pin for channel
2.
Outputs a voltage of approximately 0V
to 10V to represent 0V to the rated
output voltage.
9
CH2 EXT A MON
OUT
Output current monitor pin for channel
2.
Outputs a voltage of approximately 0V
to 10V to represent 0A to the rated
output current.
10
CH2 COM
COM pin for channel 2. Connected with
the negative output terminal of channel
2.
11
CH2 EXT ALARM
IN
Causes the alarm state to be triggered
when short-circuited with the CH2
COM pin.
12
CH2 EXT OUTPUT
IN
This pin turns the CH2 output on when
it is shorted to the CH2 COM pin.