EXTERNAL CONTROL
137
(SRE) is set 1 at this time, bit 6 (MSS) of the status
byte register is set to 1 and a service request (SRQ)
is generated.
Bit 5
ESB
Standard event status register summary
Bit 4
MAV
Message available
Bit 1
LERR
Local bus execution error
Bit 0
TMO
Local bus time-out
NOTE
An SRQ is issued through the GPIB interface only.
Reading Data from the Status Byte Register and Clearing the
Status Byte Register
Introduction
The content of the status byte register may be read
out from the controller in the following two ways:
Executing the the *STB? query.
When the *STB? query is executed the status of
the MSS bit is returned. No bits of the status
byte are cleared after reading it.
Serial polling.
When serial polling is executed, the status of the
RQS bit (bit 6) is returned. The RQS bit is only
cleared after reading it. It is impossible to read the
MSS bit. It is impossible to clear the status byte
register directly; to clear the status byte register,
you must first clear the associated event registers
that cause the bits of the status byte register to be
set.
When ESB occurs.
When an inquiry is made with the *ESR? query,
the data from the standard event register are
read and all the bits are cleared.
When the *CLS command is received.