PSB-2000 Series User Manual
120
External Control Setting (:EXT:MOD)
This command sets the external control mode.
Setting
:EXT:MOD <value>
<value> is as shown below:
0: Settings are controlled from the front panel or via
remote control.
1: External voltage control
2: External resistance control
Application
example
:EXT:MOD 1
Selects the external voltage control.
Query
:EXT:MOD?
Inquires about the setting status.
Response
example
1
Indicates that external voltage control has been enabled.
The query return has the same meaning as the setting
<value>.
Remark
This command causes an error if a sequence operation is
running.
External Control of the Voltage Output ON/OFF
(:EXT:VOLT)
This command chooses whether the external control is used to
control the output voltage or if the front panel controls are used to
control the output voltage.
Setting
:EXT:VOLT <value>
Activates or deactivates external control of the output
voltage of a single channel model.
:EXT:VOLT:A <value>
Activates or deactivates external control of the output
voltage of CH1.
:EXT:VOLT:B <value>
Activates or deactivates external control of the voltage
output of CH2.
<value> is as shown below:
0: Deactivates external control of the voltage output
(Allows the voltage output to be controlled from the front
panel or from remote control).
1: Activates external control of the voltage output.
Application
example
:EXT:VOLT 1
Selects external control of the voltage output.
Query
:EXT:VOLT?