OPERATION
71
menu and return to normal operation
. ―Ab‖ is
displayed on the MENU LED when the
tracking function is activated.
Example: Tracking function is activated.
Master-slave function
Menu Item: 10
The master-slave function allows the PSB series to
operate in parallel or series (PSB-L only). For
series operation a maximum of 2 PSB-L units can
be used. For parallel operation, a maximum of 4
(PSB-L) units or 2 units (PSB-H) units can be used.
Note that for series operation, the same PSB-L
models types must be used.
For a slave unit that is connected in series, only
the output voltages are displayed when the
output is on.
No panel operations can be performed on the
slave units. Only the Off Timer, Sequence and
External Control function settings are initialized
on the slave units.
For parallel connections ―
---
― is displayed on the
display for slave units. The monitor values are
displayed on the master unit.
CAUTION
The Hi-
Ω function is not available for master
-
slave operation.