GPM-8310 User Manual
40
Setting up data update rate
Steps
1.
Press
Setup
button.
2.
Press
Enter
button.
3.
Press down arrow key to move
cursor to the
Data Update Rate
field.
4.
Use soft keys to select and confirm
the desired option. Press the
“
More
” soft keys to toggle among
pages for further options.
Option
0.1s/0.25s/
0.5s/1s/2s/
5s/10s/20s
Measured value is updated in
accordance with the designated time
interval. The Update 5s status icon, for
example, on the display lights up in
green when 5s option is selected.
Auto
Data is only updated when a set
period (Time Out) of the input
waveform is detected.
Default value
0.25s
Steps
5.
When Auto is selected, press right
arrow key to move cursor to
Time
Out
field.