GPM-8310 User Manual
176
[:INPut]:MODE
Description
Sets or returns the voltage and current
measurement mode.
Syntax
[:INPut]:MODE {DC|AC/RMS|ACDC| VMEan}
Query Syntax
[:INPut]:MODE?
Parameter/
Return parameter
DC
AC/RMS
ACDC
VMEan
Select the dc measurement mode.
Select the ac measurement mode.
Select the acdc measurement mode.
Select the vmean measurement mode.
Example
:INPUT:MODE DC
:INPUT:MODE?
->:INPUT:MODE DC
[:INPut]:VOLTage
Description
Returns all voltage measurement settings.
Query Syntax
[:INPut]:VOLTage?
Return parameter <String>
[:INPut]:VOLTage:RANGe
Description
Sets or returns the voltage range.
Syntax
[:INPut]:VOLTage:RANGe {<Voltage>}
Query Syntax
[:INPut]:VOLTage:RANGe?
Parameter/
Return parameter
<Voltage>
15, 30, 60, 150, 300, 600(V) when the
crest factor is set to 3.
7.5, 15, 30, 75, 150, 300(V) when the
crest factor is set to 6 or 6A.
Set
Query
Query
Set
Query