GPM-8310 User Manual
126
Manual (Manual
range mode)
You can set which measurement function values
result in a D/A output of -5V, and which result in a
D/A output of +5V. By doing so, you can enlarge or
reduce (zoom) the D/A output of each channel.
For example, if you are measuring a current that
fluctuates between 0.4A and 0.6A with a measurement
range of 1A, when the D/A output range mode is
Fixed, the D/A output voltage will fluctuate between
2.0V and 3.0V. When you want to observe the
fluctuations more closely, you can use the D/A zoom
feature. If you set the D/A output range mode to
Manual and set the minimum value to 0.4 and the
maximum value to 0.6, the instrument will produce –
5V when the measured current value is 0.4A and +5V
when the measured current value is 0.6A.
Maximum and
Minimum Values
in Manual Range
Mode
The value maximum and minimum is between -
9.999E+12 and 9.999E+12.
The default value is 100.0 and -100.0.