COMMAND OVERVIEW
171
Parameter
<Boolean>0
<Boolean>1
OFF
ON
Return parameter 0
1
Turn the harmionic display off.
Turn the harmionic display on.
Example
: HARMONICS:DISPLAY:STATE OFF
: HARMONICS:DISPLAY:STATE?
->:HARMONICS:DISPLAY:STATE 0
:HARMonics:DISPlay:ORDer
Description
Sets or returns the harmonic order of the harmonic
component that is shown in graph->hormoics->bar
page for the harmonic measurement data display.
Syntax
:HARMonics:DISPlay:ORDer {<NR1>}
Query Syntax
:HARMonics:DISPlay:ORDer?
Parameter/
Return parameter
<NR1>
1 to 50 (harmonic order).
Example
:HARMONICS:DISPLAY:ORDER 1
:HARMONICS:DISPLAY:ORDER?
->:HARMONICS:DISPLAY:ORDER 1
:HARMonics:PLLSource
Description
Sets or returns the PLL source.
Syntax
:HARMonics:PLLSource {U1|I1}
Query Syntax
:HARMonics:PLLSource?
Parameter/
Return parameter
U1
Select pll source at voltage.
I1
Select pll source at current.
Example
:HARMONICS:PLLSOURCE U1
:HARMONICS:PLLSOURCE?
->:HARMONICS:PLLSOURCE U1
Set
Query
Set
Query
Summary of Contents for GPM-8310
Page 1: ...Digital Power Meter GPM 8310 USER MANUAL Rev A ISO 9001 CERTIFIED MANUFACTURER ...
Page 74: ...GPM 8310 User Manual 74 3 Press Config soft key to Enter SYSTEM CONFIG setting screen ...
Page 221: ...COMMAND OVERVIEW 221 Example STATUS QMESSAGE ON STATUS QMESSAGE STATUS QMESSAGE 1 ...
Page 239: ...APPENDIX 239 Status system The diagram below is a description of the status system ...
Page 243: ...APPENDIX 243 Dimensions Units mm ...