GPM-8310 User Manual
206
:NUMeric:LIST:VALue
Description
Returns the harmonic measurement numeric list
data.
Query Syntax
:NUMeric:LIST:VALue? {<NRf>}
Parameter
<NRf>
1 to 8 (item number)
Example
• if <NRf> is specified :NUMERIC:LIST:VALUE? 1
-> 103.58E+00,NAN,103.53E+00,0.09E+00,2.07E+00,
0.04E+00,.. (omitted)..,0.01E+00,0.01E+00
(up to 52 data values)
• if <NRf> is omitted (when :NUMeric:LIST:NUMber
is set to 5)
:NUMERIC:LIST:VALUE?
-> 103.58E+00,NAN,103.53E+00,0.09E+00,2.07E+00,
0.04E+00,..(omitted)..,0.00E+00,0.00E+00
(up to 52*5 = 260 data values)
• When :NUMeric:FORMat is set to {FLOat}
:NUMERIC:LIST:VALUE?
-> #N (N-digit byte number)(data byte sequence)
NOTE
A single numeric list data item consists of up to 52
items of numeric data in the following order:
TOTal, DC, 1st
harmonic, ..., :NUMeric:LIST:ORDer.
If <NRf> is specified, only the numeric list data of
the specified item number is output (up to 52
items of data)
If <NRf> is omitted, the numeric list data of item
numbers from 1 to :NUMeric:LIST:NUMber is
output in order (up to 52 times the number
specified by :NUMeric:LIST:ORDer)
:NUMeric:LIST:NUMber
Query
Set
Query
Summary of Contents for GPM-8310
Page 1: ...Digital Power Meter GPM 8310 USER MANUAL Rev A ISO 9001 CERTIFIED MANUFACTURER ...
Page 74: ...GPM 8310 User Manual 74 3 Press Config soft key to Enter SYSTEM CONFIG setting screen ...
Page 221: ...COMMAND OVERVIEW 221 Example STATUS QMESSAGE ON STATUS QMESSAGE STATUS QMESSAGE 1 ...
Page 239: ...APPENDIX 239 Status system The diagram below is a description of the status system ...
Page 243: ...APPENDIX 243 Dimensions Units mm ...