Reverse osmosis system
GENO-OSMO-X
Order no.
134 750 927-inter Edited by: phar/mrie G:\BA-750927-inter_134_GENO-OSMO_X.docx
61
2 | How to remedy warnings
Table G-1: Warnings
This is what you observe
This is the cause
This is what to do
W006
Malfunction of parameter ad-
justment EH
(water softener)
Malfunction of parameter adjust-
ment EH (water softener).
Different software versions of
OSMO-X and water softener. Per-
form an update.
Incorrect water softener selected.
W007
Firmware of water softener out-
of-date
Firmware of water softener out-of-
date.
Software version of operating ele-
ment not compatible with basic cir-
cuit board. Update basic circuit
board.
W008
Malfunction of parameter ad-
justment DH (pressure booster)
Malfunction of parameter adjust-
ment DH (pressure booster).
Different software versions of
OSMO-X and pressure booster sys-
tem.
Perform an update.
Incorrect pressure booster system
selected.
W009
Firmware of pressure booster
out-of-date
Firmware of pressure booster out-
of-date.
Software version of operating ele-
ment not compatible with basic cir-
cuit board. Update basic circuit
board.
W010
HMS bus module not sup-
ported.
HMS bus module not supported.
Module not recognized properly or a
module type that is not yet being
supported was plugged in.
W021
LF permeate conductivity
RO1CQ1
The conductivity monitoring of the
permeate is programmed to “sig-
nal“ and the conductivity continu-
ously exceeded the threshold
value for longer than the pro-
grammed delay time.
Typical system behaviour:
Following the start of the permeate
production, the permeate conduc-
tivity quickly and within minutes
drops (first permeate) until a more
or less constant value is reached.
A non-functioning pre-treatment,
adversely affects (scaling) this con-
stant value.
Also refer to chapter F-4.1:
Conductivity limit value too low or
delay conductivity signal/error pro-
grammed too short.
If necessary notify Grünbeck’s tech-
nical customer service/authorised
service.
W022
DOS low RO1CL1
Only in case of DOS pre-treatment:
Dosing agent will be used up soon.
Insert new tank with dosing agent
and confirm in subsystem menu, so
that the range estimate is restarted.
W023
DOS low RO1CL3