Reverse osmosis system
GENO-OSMO-X
Order no. 134 750 927-inter Edited by: phar/mrie G:\BA-750927-inter_134_GENO-OSMO_X.docx
24
D Installation
1 | General Installation Instructions
The installation site must provide adequate space. Only the left
side (pressure pipes) of the system may be placed flush to the
wall. In front, behind and on the right side of the system, adequate
clearance for installation and service work must be provided
(> 50 cm). A foundation of a sufficient size and adequate load car-
rying capacity must be provided. The required connections must
be provided prior to the installation. For dimensions and connec-
tion data, please refer to table C-1.
1.1 Preliminary work
Unpack all system components.
Check for completeness and soundness.
Place the reverse osmosis system at the designated location.
Note:
Also observe the operation manuals that have been sup-
plied with the optional accessories (see chapter C-6.2) for your
system (if applicable).
2 | Water installation
Certain binding rules must always be observed when installing a
reverse osmosis system. Additional recommendations are given
in order to facilitate the handling of the system. The installation
instructions described below are also illustrated in fig. D-1.
Binding rules
The installation of a reverse osmosis system represents a major
interference with the drinking water system. Therefore, only au-
thorised experts may install such systems.
Please observe the local installation guidelines and the gen-
eral regulations.
Install a drinking water filter upstream of the system (e.g.
BOXER K).
A system separator must be installed upstream as well.
Install a water softener upstream.
Mount an activated carbon filter, if necessary (take water
analysis into consideration).
Provide a drain connection (minimum DN 50) to discharge
the concentrate.
Note:
If the concentrate is directed to a lifting system, the delivery
rate of the lifting system should at least have the concentrate vol-
ume flow indicated in table C-1.
Warning!
The installation site must have a floor drain. If no floor
drain is available, an adequate safety device needs to be in-
stalled (refer to option C-9).
Warning!
In case of a power failure, floor drains with a dis-
charge to a lifting system will be out of operation.
Summary of Contents for GENO-OSMO-X
Page 2: ......