PART THREE - OPERATION
SECTION 5 - TEST/MAINTENANCE
PRO-series Model C3 Conductivity Transmitter
Rev. 0-201
60
☞
5.4 Firmware (EPROM
VERSION) Checking
5.5 SELECT SIM
Measurement
2.
Press
ENTER key
to display a screen like
.
NOTE:
The mA output test signal is now active. Its
value is shown on this screen.
3.
Use
arrow keys
to adjust the displayed value to obtain
the desired mA test signal.
4.
To remove the output test signal and return to the pre-
vious level of the TEST/MAINT menu branch, press
ESC key
or
ENTER key
.
You can check the firmware EPROM version used in the
transmitter.
1.
With the
screen displayed, press
Ø
Ø
key
until
screen appears.
2.
Press
ENTER key
to view the EPROM version screen.
3.
To return to the previous level of the TEST/MAINT
menu branch, press
ESC key
or
ENTER key
.
You can simulate a measured value to make the analog
output respond accordingly. First, select the type of simu-
lated value using this subsection. Then, set the desired
simulation value following the steps in subsection 5.6.
1.
With the
screen displayed, press
Ø
Ø
key
until
screen appears.
2.
Press
ENTER key
to display a screen like
. Use
Ø
Ø
and
×
×
keys
to view both
choices: