PART THREE - OPERATION
SECTION 3 - TRANSMITTER CONFIGURATION
PRO-series Model C3 Conductivity Transmitter
Rev. 0-201
44
3.4 Configuring
Analog Output
SET PARAMETER
(representation)
SET 4 mA and
20 mA VALUES
(range expand)
The transmitter provides an isolated 4-20 mA analog output.
During normal measurement operation, the output is active
but can be held at the last measured value for up to 30 min-
utes by using the “HOLD OUTPUT” function in the
TEST/MAINT menu. (See PART THREE, Section 5.2 for
details.) During calibration, the output is automatically held
at the last measured value and, upon completion, returned
to its active state.
The output can be assigned to represent the SENSOR
(measured conductivity, resistivity or TDS) or measured
TEMPERATURE.
1.
With the
screen displayed, press
×
×
key
-- not
Ø
Ø
key
-- twice to display
.
2.
Press
ENTER key
to display
.
3.
Press
ENTER key
again to display
.
Use
Ø
Ø
and
×
×
keys
to view both choices. With the de-
sired choice displayed, press
ENTER key
to enter this
selection.
Parameter values can be set to define the endpoints at
which the 4 mA and 20 mA analog output values are desired.
1.
With the
screen displayed, press
Ø
Ø
key
once to display
.
2.
Press
ENTER key
to display a screen like
. Use
arrow keys
to set the dis-
played value at which 4 mA is desired, and press
ENTER key
to enter the value.
3.
After the
screen re-appears, press
Ø
Ø
key
once to display
.
4.
Press
ENTER key
to display a screen like