Rev. 0-201
PRO-series Model C3 Conductivity Transmitter
43
3.3 SET
°
C OR
°
F
(temperature display
format)
•
Shortened Sensor Cable: To increase the transmitter tem-
perature reading to match the known solution temperature,
decrease the T FACTOR by 3.85 ohms for each
°
C difference.
•
Added Interconnect Cable: To decrease the transmitter tem-
perature reading to match the known solution temperature,
increase the T FACTOR by 3.85 ohms for each
°
C difference.
Example:
Suppose the known solution temperature is 50
°
C
and the transmitter reads 53
°
C due to interconnect cable re-
sistance. Multiply the 3
°
C difference by 3.85 ohms to get 11.55.
Then increase the sensor T FACTOR by adding 11.55 to it and
entering that value. If, due to a shortened sensor cable, the
transmitter was reading 3
°
C less than the known solution tem-
perature you would decrease the sensor T FACTOR by
subtracting 11.55 from it.
*Sensors with a Pt 100 temperature element provide inherently less
accurate temperature readings and are not recommended.
3.
After the
screen re-appears, press
ESC key
twice to return to the
screen.
The MEASURE screen can be set to display temperature
values in
°
C or
°
F. In either case, the display resolution for
measured temperature is always “XX.X.”
1.
With the
screen displayed, press
×
×
key
-- not
Ø
Ø
key
-- twice to display
.
2.
Press
ENTER key
to display a screen like
. Use
Ø
Ø
and
×
×
keys
to view both
choices (
°
C or
°
F). With the desired choice displayed,
press
ENTER key
to enter this selection.