PART THREE - OPERATION
SECTION 1 - USER INTERFACE
PRO-series Model C3 Conductivity Transmitter
Rev. 0-201
30
1.4 MEASURE Screen
(normal display mode)
☞
The MEASURE screen is normally displayed. Pressing the
MENU key
temporarily replaces the MEASURE screen with
the top-level “MAIN MENU CALIBRATE” branch selection
screen. Using the keypad, you can then display other
screens to calibrate, configure or test the transmitter.
If the
keypad is not used within 30 minutes, except during
calibration or while using specific transmitter test/
maintenance functions, the display will automatically
return to the MEASURE screen.
To display the MEASURE
screen at any time, press the
MENU key
once and then
press the
ESC key
once.
The MEASURE screen can show four different readout ver-
sions. To select between them, in continuous loop
sequence, press the
Õ
Õ
or
Ö
Ö
key
. These are examples of
the different versions:
Ö
Ö
Ö
Ö
NOTE:
When the transmitter returns to its normal
MEASURE screen mode, the appearing readout is
always the version last selected. Note that three
MEASURE screen readout examples show the
factory-default “COND” notation on their top lines,
illustrating the transmitter notation feature. To cre-
ate your own notation, refer to PART THREE,
Section 3.2, subheading “ENTER NOTE (top line of
MEASURE screen).”
When the measured value is beyond the transmitter meas-
uring range, a series of “ + ” or “ - ” screen symbols appear,
respectively indicating that the value is above or below
range.