PART THREE - OPERATION
SECTION 3 - TRANSMITTER CONFIGURATION
PRO-series Model C3 Conductivity Transmitter
Rev. 0-201
40
SELECT CELL K
(sensor’s GLI-certified
“K” value)
A.
Starting with extreme left character position, use
×
×
and
Ø
Ø
keys
to select the desired first character.
B.
Press
Ö
Ö
key
once to select the next character, and
use
×
×
and
Ø
Ø
keys
to select its desired character.
C. Repeat procedure until desired notation is displayed.
3.
Press
ENTER key
to enter the displayed notation.
Each GLI sensor has a unique, certified “K” value shown on
a label attached to the sensor cable or to the inside cover of
its optional junction box head. By entering this “K” value,
calibration (including zeroing) is only necessary when this
sensor is replaced. This also sets the transmitter measuring
range to match the inherent range of the sensor’s cell con-
stant.
1.
With the
screen displayed, press
Ø
Ø
key
once to display
.
2.
Press
ENTER key
to display
.
3.
Press
ENTER key
again to display a screen like
. Use
Ø
Ø
and
×
×
keys
to select the
nominal cell category that corresponds to the sensor’s
GLI-certified “K” value, and press
ENTER key
to enter
this selection.
4.
After the
screen re-appears, press
Ø
Ø
key
once to display
.
5.
Press
ENTER key
to display a screen like
. Use
arrow keys
to adjust the
displayed value to exactly match the sensor’s GLI-
certified “K” value, and press
ENTER key
to enter the
value.