PART THREE - OPERATION
SECTION 4 - TRANSMITTER CALIBRATION
PRO-series Model C3 Conductivity Transmitter
Rev. 0-201
52
SET T FACTOR
(sensor’s GLI-certified
“T” factor)
6.
Press
ENTER key
to display
.
7.
Press
ENTER key
again to display a cell category selec-
tion screen like
. Use
×
×
and
Ø
Ø
keys
to select the nominal cell category that corre-
sponds to the sensor’s GLI-certified “K” value (shown
on label attached to sensor cable or to inside cover of
its optional junction box). Then press
ENTER key
to
enter this selection.
8.
After the
screen re-appears, press
Ø
Ø
key
once to display
.
9.
Press
ENTER key
to display a cell “K” value screen like
. Use
arrow keys
to adjust the
displayed value to exactly match the sensor’s GLI-
certified “K” value, and press
ENTER key
to enter the
value.
After entering the “K” value, complete DRY-CAL calibration
by entering the GLI-certified “T” factor:
1.
With the
screen displayed, press
ESC key
once to display
.
2.
Press
Ø
Ø
key
once to display
.
3.
Press
ENTER key
to display
.
4.
Press
Ø
Ø
key
once to display
.
5.
Press
ENTER key
to display a “T” factor value screen
like
. Use
arrow keys
to adjust the
displayed value to exactly match the sensor’s GLI-
certified “T” factor, and press
ENTER key
to enter the
value.