-25-
for neonates:
preset range: 70~100mmHg, default value: “90” mmHg;
for infants:
preset range: 80~130 mmHg, default value: “110” mmHg;
for adults
: preset range: 80~180mmHg, default value: “140” mmHg.
If the pressure drops down slowly under 10mmHg compared with the preset value due to
little air leakage in the pneumatic system when time passes by, the monitor will re-inflate
to maintain the cuff pressure close to the preset pressure value.
Note:
the unit of cuff pressure is the same with the NIBP unit in NIBP Setup.
“Duration”:
After presetting the cuff pressure, you need to set the time period for
maintaining the preset pressure after inflation. “5, 6, 7,…120” minutes adjustable. The
default value is “40” minutes.
If the set value is “xx” minutes, the monitor will count down from “xx” minutes
automatically when starting cuff inflation. When time is up, it will deflate automatically.
“Alert T”:
the alert time for reminding user that the operation of tourniquet is going to be
end after this time period. 1 to 60 minutes adjusting range with 1 minute step, the default
value is “5” minutes. If the set value is “xx” minutes, the monitor will produce alarm sound
until ending deflation when counting down time reaches to “xx” minutes. The alarm type is
high priority alarm. (For example: the duration is 40 minutes, the alert time is 5 minutes, the
alarm will ring for prompt when the duration counting down to 5 minutes. The Prompt Info
area starts to prompt: TOUR C-D 300 seconds. )
“Start”:
shift cursor to “Start” and press “ ” key, “Start” becomes “Stop” and meanwhile
the blood cuff starts being inflated; Pressing “Stop” button can stop using this function.
After deflation, it will change to “Start” again.
4.5.4 Nurse Call
Figure 4.11 Nurse Call Setup Screen
Screen Description:
“Output level”:
two options “low” or “high” output levels are available.
When the calling system in hospital works in “Normal Open” mode, “low level” should be
selected.
When the calling system in hospital works in “Normal Close” mode, “high level” should be
selected