-26-
“Source”:
three kinds of alarm sources can trig the nurse call: high level alarm, medium level
alarm and low level alarm (multi-optional). If you don’t make choice, nurse call signal will not
be sent out.
“Duration”:
two options “pulse” or “continuous” output modes are available;
“continuous”:
the continuous mode of output means the nurse call signal will keep until
the selected alarm source(es) disappear, i.e. the signal will last from starting alarm to
stopping alarm.
“pulse”:
the output nurse call signal is
pulse signal which lasts for 1 second. When several
alarms occur at the same time, only one pulse signal will be sent out.
Note:
Nurse Call function can not be regarded as main alarm notice method, please do not
entirely relay on it. You should combine parameter values with alarm level and
patient’s clinical behavior and symptom to determine patient’s status.
4.5.5 System Setup
Figure 4.12 System Setup Screen
Screen Description:
“Vol”:
set beeper volume, “0~7” level adjustable, the set “0” i.e. no sound.
“key beep”:
to turn on/off key beep;
“LANG”:
language selection. “ENG” for English.
“priority”:
priority of “PR” value or “HR” value display. The default set is “HR”.
“backlite”:
turn on/off backlight;
“contrast”:
adjust LCD display contrast, “0~31” level adjustable;
“care mode”:
“Demo” shows the demo waveforms and data. In the demo state, all the
signals and data are generated from the monitor for demonstration and testing purpose.
“Real” shows the real time waveform, i.e. normal monitoring status;
BT SD:
turn on/off the pulse beeping sound.