-5-
2
SYS:
display systolic pressure value
3
DIA:
display diastolic pressure value.
4
MAP:
Display mean arterial pressure or measuring time of the latest group of NIBP
measurement; they will be displayed alternately. The format of NIBP measuring time is
“hh:mm”. If the tourniquet is in use, the cuff pressure will be diplayed here.
Note:
two formats to display NIBP value: “×××mmHg” and “××.×kPa”. Refer to section “4.4.2
NIBP Setup” to set the unit of NIBP value; the conversion relation between “mmHg” and
“kPa”: 1mmHg=0.133kPa.
5
HR(priority indicator)
: if HR indicator is on, it indicates that the numerical value beside is
HR measuring value;
6
Display HR or PR value:
when the set of “Setup Menu System priority” is “HR”, it
shows HR value here preferentially; if the set is “PR”, PR value will be shown preferentially.
7
PR(priority indicator):
if PR indicator is on, it indicates that the numerical value beside is
pulse rate value; Unit: “bpm (beats per minute)”.
8
SpO
2
:
Display SpO
2
value; Unit: “%”
9
“
”:
Bar-graph of pulse intensity
10
LCD
panel
11
Pulse sync indicatoratient category indicator:
“
” for adult; “ ” for pediatric; “
” for
neonate; Patient category is selected under sub-menu “Patient Info” within the setup
menu.
12
Pulse
sync indicator
: Cardio-pulse/pulse sync indicator. When
HR priority indicator is on,
its flashing is synchronized with heart beat; When
PR priority indicator is on, its flashing is
synchronized with pulse.
13
NIBP:
start/cancel NIBP measurement
14
Alarm silence key:
Enable/disable alarm silence function. When the alarm silence
indicator on the left of keys is on, it means the system is in alarm silence status and it lasts
this status for 2 minutes. When finishing counting down, the system will resume normal
alarm status automatically, if alarm event occurs at this time the alarm sound will be
effective again.
15
Alarm silence indicator:
When it is on, it indicates that the monitor stays in alarm silence
status.
16
Print:
the internal printer is optional, press this key to print the current measuring
data;
17
Up: shift cursor forward/upward
18
OK: to confirm selection or modification