-22-
There are 8 functional groups for setting parameters: “ECG TEMP, SpO
2
, NIBP, Nurse Call, System,
Patient Info, Date/Time and Default” on the Setup Menu Screen.
1. Press“
” key or “
” key to shift cursor to corresponding functional group setting.
2. Pres “
” key to confirm and enter into corresponding functional parameter setup screen.
3. Pres “
” key under the setup menu will print ECG waveform.
4. Press “
” to exist from Setup Menu Screen.
At Setup Menu Screen or its submenu screen, when pressing “
” key,
the default display
screen will be printed.
The following will cover each functional parameter’s setting up.
4.5.1 ECG and Temperature Setup
Figure 4.8 ECG/TEMP Setup Screen
Screen Description:
“Lead”:
ECG lead selection:
,
,
, AVR, AVL, AVF or V;
“Gain”:
ECG waveform scale:
“×1/2”-
waveform reduced to half of nominal scale
“×1”-
nominal waveform scale
“×2”-
waveform with doubled scale
“1mV”:
generating internal 1mV calibration signal. This signal is used to test the function of
the machine. It is not used during normal operation. The default set is off.
“HR
”:
HR
alarm switch; “
” indicates HR
alarm is on; “
” indicates HR alarm is off.
“HR Hi/Lo”:
high/low limit of HR
alarm;
“TEMP
”:
temperature
alarm switch; “
” indicates temperature
alarm is on; “
”
indicates temperature alarm is off.
“TEMP Hi/Lo”:
high/low limit of temperature alarm;
“Unit”:
body temperature unit. Two options: “
” or “
”. Conversion relation: 1
= (
X1.8)
+32.
“T probe”:
the type of temperature probe “KRK”;