12. Frequency response: 0.5 Hz
40Hz(
,
)
5.2 TEMP Monitoring
1. TEMP measuring range: 25.0~45.0
2. TEMP measuring accuracy: ±0.2
3. TEMP responding time: ≤150s
5.3 NIBP Monitoring
1. Measuring method: Oscillometric Technique
2. Pneumatic pressure measuring range: 0 mmHg~300mmHg
3. Accuracy of pressure measurement: ±3 mmHg
4. Cuff inflation time: <10 seconds (typical adult cuff)
5. Measurement time on the average: < 90 seconds
6. Air release time while the measurement is canceled: <2 seconds (typical adult cuff)
7. Initial cuff inflation pressure
Adult: <180 mmHg; Infant: <120 mmHg; Neonate: <90 mmHg
8. Overpressure protection limit
Adult: 300 mmHg; Infan: 240mmHg; Neonate: 150 mmHg
9. NIBP measurement range:
press
unit
Adult Infant
Neonate
SYS mmHg
40~255
40~200
40~135
MAP mmHg
20~215
20~165
20~110
DIA mmHg
10~195
10~150
10~95
10. NIBP accuracy:
Maximum mean error: ±5 mmHg
Maximum Standard deviation: 8 mmHg
11. Measurement mode: Manual, Auto, STAT
5.4 SpO
2
Monitoring
1. Transducer: dual-wavelength LED
2. SpO
2
measuring range: 35%~100%
3. Low perfusion capability: 0.4%~5%
4. SpO
2
measuring accuracy:
50%~100% 3%
*NOTE: Accuracy of SpO
2
should be the root-mean-square(rms) of difference.
5.5 Pulse Monitoring
1. Pulse rate measuring range: 30bpm~240bpm
2. Pulse rate measurement accuracy: ±2bpm or ±2%, whichever is greater.