Distributed by
Rescue Life - User Manual Rev. 3.1
79
NIBP (optional)
•
Technique:
oscillometric.
•
NIBP Accuracy:
Meets ANSI/AAMI SP10-2002, EN 1060-4.
•
Patient Application:
Adult/Paediatric/Neonatal.
•
Systolic Range:
Adult: 40-260mmHG, Pediatric: 40-160 mmHG,
Neonatal: 40-130 mmHG.
•
Range MAP:
Adult: 26-220 mmHG, Paediatric: 26-133
mmHG, Neonatal: 26-110 mmHG.
•
Diastolic Range:
Adult: 20-200 mmHG, Paediatric: 20-120 mmHG,
Neonatal: 20-100 mmHG.
•
PR RANGE:
30-220 BPM.
•
PR Accuracy:
+/-2% or +/3 bpm.
•
Transducer Accuracy:
+/- 3 mmHg over full range in operating conditions.
•
Operating Modes:
Manual, Long Term Automatic, Stat, Service.
•
Auto Interval Periods:
1,2,3,4,5,10,15,30,60 and 90 minutes.
•
Safety & Regulatory Standards:
IEC60601-1, IEC/EN60601-2-30, AAMI SP10, OIML R 16-2.
EN1060-1, EN1060-3, EN1060-4.
External pacemaker (optional)
•
Type: rectangular wave.
•
Operating mode: Fixed, on demand.
•
Pulse rate: 30 bpm to 250 bpm, adjustable in steps of 5 bpm.
•
Impulse duration: 22.5 ms.
•
Pulse current: 0 to 150 mA, adjustable in steps of 5 mA.
•
Amplitude: max 150 V.