Rescue Life - User Manual Rev. 3.1
62
Switch off defibrillator by pressing On/Off key.
On the graphic hardcopy if a defibrillation shock was recorded the graph will show the
base line for about 1 second (shock event).
DATE AND TIME SETUP
To set-up the real time clock, switch on the RESCUE LIFE with paddles connector
disconnected.
On the screen will be displayed the battery status, the date and time. The
SETCLK
-
F4
key enables the clock set-up. With the Speed Dial select the value to change and
rotating it the value can be modified. Press the Speed Dial once the value has been
changed.
To exit the clock set-up press again the
SETCLK -
F4
key.
PRINTING AND PAPER CHANGE
The printing can be done in manual mode or automatic mode.
To select print mode enter the menu and select MANUAL or AUTOMTIC mode from PRINT
MODE.
MANUAL MODE
To start printing press the
- F2
key. The hardcopy will start with the set-up
parameters and with the ECG trace. The hardcopy will keep the LCD display set-up
(amount of the traces, group, speed the gain)
To stop the printing, press again the
–
F2
key.
If SpO
2
sensor is connected (optional) in hardcopy user will found value of oxygen
saturation.
AUTOMATIC MODE
RESCUE LIFE will print an ECG frame of 6 seconds when the charge starts.
When the shock is released the energy delivered, as well as the time stamp will be
printed together with a 6 seconds after shock ECG frame.
The manual printi
ng (using F2 key, “PRINT”) is
operational independently of the printing
mode.
Do not leave the device without supervise during printing. The thermal printer can be
damage by a prolonged use.
When the paper is finished the green light on the printer cover button will switch on.