IMPENDANCE LIMITS
RESCUE LIFE does not release the shock if the patient impedance is less than 25 ohm
or over 200 ohm.
SYNC/NO SYNC MODE
When the
RESCUE LIFE
is on it is automatically set on no-sync mode.
Only the operator can set the sync/no-sync mode which is clearly displayed on the
screen. The device can not automatically set the sync mode.
In the sync mode the device
releases the defibrillation shock only when the ‘R’ peak in
the ‘QRS’ complex is detected.
The maximum response time between the “R” peak and the def
ibrillation shock is less
than 60ms.
CHARGING TIME TO ACHIEVE THE MAXIMUM ENERGY (230 JOULE)
When the device is connected to the AC supply (nominal AC voltage) and batteries are
100% charged :< 8 sec.
When the device is connected to the AC supply (AC voltage 90%) after 15 shocks :<
10 sec.