Distributed by
Rescue Life - User Manual Rev. 3.1
23
OPERATIONAL SCREEN FEATURES
In the Operational screen are shown all vital parameter and message as well as function
button and some setting.
MESSAGGE ZONE
: area where messages are shown.
ECG TRACE
: area where ECG trace is shown. In menu it is possible to choose traces to
display, signal gain and speed.
BATTERY STATUS INDICATOR
: battery status is shown. If charge value is under 40
% indicator is yellow and if value is under 10 % indicator is red. Connecting power cord
indicator becomes green.
When the battery value is under 10% a message appears on display. Insert the power
supply cord in RESCUE LIFE socket, otherwise system will shout down automatically in
2 minutes. During this time defibrillation functions are disabled.
DATE AND TIME
: date and time are shown.
OPERATION MODE
: operation mode selected is shown. It is possible to select
MANUAL, ADVISORY (ADV) and SEMIATUMATIC (AED) mode.
ALARMI ZONE:
area where alarm are shown.
HEART RATE
: heart rate is shown. It is possible to activate or deactivate acoustic beep
of heartbeat.
ENERGY SELECTION
: energy set is shown. To set energy turn knob clockwise to
increase energy and counterclockwise to decrease energy.
CHARGE/DISCHARGE INFORMATION
: when charging starts the status bar indicates
the progress of the charge.
It also indicates when discharge is done and the amount of discharge when the charge
level of the battery is under 15 %.
SYNC STATUS
: it is shown if the sync mode is active or not.