BATTERY CHARGE
When the message of the battery status displays a value under 25%, batteries should
be charged. Insert the power supply cord in the RESCUE LIFE socket (located on the
back side) and connect to the AC line. The battery status led will switch on.
When the charge finished the led will switch off. To see the battery charge status switch
on the device with pads connector not attached.
When the battery value is under 10% a message appears on display
: “CHARGE BATTE
RY
OR SYSTEM
WILL SHUT OFF”
. Insert the power supply cord in RESCUE LIFE socket,
otherwise system will shout down automatically in 2 minutes.
During this time defibrillation functions are disabled.
W
HEN THE DEVICE IS OFF
,
DO NOT LEAVE THE
AC
CHARGER CONNECTED MORE THAN
3
HOURS
.
I
F AFTER
THIS TIME THE CHARGING LIGHT DOES NOT GO OFF
,
PLEASE CONTACT THE SERVICE CENTER FOR
CHANGING THE BATTERIES
.