HOW TO PREPARE THE PATIENT
The connection of the electrodes is possible in two ways, depending on what you want:
•
ANTERIOR-POSTERIOR Connection (only pacing).
Front pad positioned on the left part of the chest, halfway between the xiphoid
apophysis and the left nipple at the sites V2-V3.
Back pad positioned immediately below the left shoulder blade side of the spine
at the same level than the front.
•
ANTERIOR-ANTERIOR connection - (for defibrillation and pacing).
Front pad positioned below the right clavicle.
Lateral plate positioned on the middle axillary line at the height of his left
nipple, position V6.
PACING PROCEDURE
1.
Switch on the device pressing the
ON/OFF
button. Connect and lock the disposable
pads plug to start the operation.
2.
Place the disposable pads on the patient then press PACER – F3 key.
3.
Press F1 then use the Speed Dial to set the value of the pacing rhythm, then press
F1 key again.
4.
Press F2 then use the Speed Dial to set the value of the pacing current, then press
F2 key again.
5.
Press F3 to select pacing mode (default mode is fixed).
6.
Press F4 key to start pacing, then press F4 key again to stop pacing.
7.
Press F5 key to exit and return to operational screen.