APPENDIX D
TECHNICAL FEATURES
ECG Monitoring
• Patient connection
:
Defibrillation pads and 5 or 10 wire ECG cable.
• Bandwidth:
0.6 to 40 Hz (-3 dB) in monitor mode.
0.05 to 120 Hz (-3 dB) in diagnostic mode.
• ECG trace
parameters:
Velocity : 5,10,25, 50 mm/sec.
Gain: 2,5, 5, 10, 20, 40 mm/mV with patient cable. AUTO with pads.
Alarm: HR max settable 250 bpm and HR min 20 bpm.
Filters: 50/60 Hz, EMG filter, base line.
Traces: 3+3+1 (I,II,III
–
aVR,aVF,aVL, V) with 5 wires patient cable.
Traces: 3+3+6 (I,II,III
–
aVR,aVF,aVL
–
V1 to V6) with 10 wires patient cable.
• Heart rate:
Digital readout on the display from 20 a 300 bpm (± 5% or ±3 bpm, on higher
value).
Defibrillator
• Operation mode:
Manual, Advisory, Semiautomatic AED.
• Defibrillabile impedance:
Compensated from 25 ohm to 200 ohm.
• Energy charging time:
< 8 sec (with batteries fully charged).
•
Manual Mode:
Syncro/Asyncro.
•
Defibrillation pads:
Standard or disposable, adult and pediatric.
•
Waveform:
Biphasic Truncated Esponential (BTE) with impedance compensation.
HiCAP Technology (Large Storage Capacitor).