EC 87 Inflate Timeout, Air Leak or Loose Cuff
Corrective Action: Check that the hose is connected to the system and the cuff.
Check that the cuff is properly tightened.
Check that the cuff is in the correct position.
Check that the correct size cuff is being applied.
Check that the cuff is not leaking air.
Check that the hose connections are not damaged or loose.
EC 88 Safety Timeout
Corrective Action: Check the patient.
Check that the cuff is in the correct position.
The patient may have been moving too much.
Take another BP measurement.
EC 89 Cuff Overpressure
Corrective Action: Check that the correct size cuff is being applied.
Check that the hose has no sharp bends or is pinched.
Check that the cuff is in the correct position.
Check that the patient is not lying on the cuff.
EC HIGHER THEN 89 System error
Corrective Action: Service may be required. Call ELPRO S.r.l. representative.