Rescue Life - User Manual Rev. 3.1
25
When menu screen is open the only keys to be active are F5 and Speed Dial.
SET-UP MENU FIELDS:
DEF MODE:
select the function mode: manual (MANUAL), advisory (ADV) or
semiautomatic (AED).
ECG TRACE:
select ECG traces that the user choose to display or print. When the patient
cable is connected, users can select which groups of traces to display. When the patient
cable is disconnected, the ECG trace is acquired by defibrillation pads (lead II).
ECG MODE:
it refers to the bandwidth of the signal.
In MONITOR mode bandwidth in
0,6 Hz to 40 Hz, in DIAGNOSTIC mode bandwidth is 0,05 Hz to 120 Hz. Notch filter is
always active. If the ECG trace is acquired by defibrillation pads active mode is
MONITOR.
ECG SETUP:
select the speed of ECG traces (display and printer) and the ECG gain.
ALARMS:
to set alarms.
HR MAX:
set the maximum heart rate alarm.
HR MIN:
set the minimum heart rate alarm.
SpO
2
MIN: s
et the SpO
2
minimum alarm.
SOUND:
enable/disable all the alarms
HR BEEP:
Enable/disable the heart rate beep.
PRINT MODE:
Set the print mode automatic or manual.
NIBP SETUP:
set
NiBP parameters.
SAVE:
Save the actual settings as default and exit the menu.
EXIT:
exit the menu.