Remote Operation
Publication 31470-001, Rev C, November 2, 2017
3-79
Table 3-14: HP438 Emulation Command Set
Command
Description
RC
Recall previous instrument state
RH
2
Do a range hold
RL0
Turn off rel mode
RL1
Turn on rel mode
RM
2
Set manual range
RV
Ask for status request mask
SM
Ask for status message
ST
Store instrument state
TR0
Trigger hold mode
TR1
Trigger single measurement
TR2
Trigger full measurement with settling
TR3
Free run trigger mode
ZE
Zero sensor (precede with AE or BE)
Notes:
1.
A numeric entry is required by these GPIB codes, followed by the EN suffix.
2.
The 8650B is always able to measure over its entire dynamic range; there is no need to specify the range.
Therefore, range related commands have no effect on the measurement capability of the 8650B. The auto
range, range hold, and set range commands only offset the analog output voltage, and only in HP436,
HP437, or HP438 GPIB emulation modes. In these emulation modes (when using a single sensor, and not
measuring in a relative mode), the power will be scaled to a range of 0 to 1 volts, representing the relative
power within the current 10 dB range of the 8650B. The range hold and set range commands will simulate
locking the range of power represented by the output voltage.