Power Sensors
Publication 31470-001, Rev C, November 2, 2017
B-9
Figure B-1: 80401A Modulation-Related Uncertainty
B.2.2.1
80401A BAP Mode Limitations
The minimum input level is -40 dBm (average); the minimum pulse repetition frequency is 20 Hz. If the input
signal does not meet these minima,
NoSync
will flash on the display to the right of the line configured for BAP
mode sensor to indicate that the input is not suitable for BAP measurement. The 8650B will continue to read the
input but the BAP measurement algorithms will not be able to synchronize to the modulation of the input; the
input will be measured as if the 8650B were in MAP mode. In addition, some measurement inaccuracy will result
if the instantaneous power within the pulse falls below -43 dBm; however, this condition will not cause
NoSync
to flash.