Series 8650B Universal Power Meters
B-10
Publication 31470-001, Rev C, November 2, 2017
Figure B-2: 80601A Modulation-Related Uncertainty
B.2.2.2
80601A BAP Mode Limitations
The minimum input level is -35 dBm (average); the minimum pulse repetition frequency is 20 Hz. If the input
signal does not meet these minima,
NoSync
will flash on the display to the right of the line configured for BAP
mode sensor the front panel to indicate that the input is not suitable for BAP measurement. The 8650B will
continue to read the input but the BAP measurement algorithms will not be able to synchronize to the
modulation of the input; the input will be measured as if the 8650B were in MAP mode. In addition, some
measurement inaccuracy will result if the instantaneous power within the pulse falls below -38 dBm; however,
this condition will not cause
NoSync
to flash. See Section 2.4.2 for modulation bandwidth limitations below 200
MHz. When the modulation bandwidth is below 200 MHz, the 806XXA sensors’ performance is equal to that of
the 804XXA sensors.