Front Panel Operation
Publication 31470-001, Rev C, November 2, 2017
2-37
Like the Peak Hold feature, the Crest Factor feature also has the same minimum average power level restriction
of -20 dBm. This has been implemented to increase accuracy and maintain wide bandwidth of Crest Factor
measurement over the peak power range of -20 dBm to +20 dBm. The Crest Factor selection will also affect
Histogram measurement if there is power below -20 dBm. To obtain accurate average power measurements
below -20 dBm, remove the display line containing Crest Factor (or, in remote control, send the command which
disables the Crest Factor feature).
NOTE: [Reset Line n]
for the Crest Factor also resets Peak Hold.