Remote Operation
Publication 31470-001, Rev C, November 2, 2017
3-83
3.5.2
Manual Averaging
The averaging number can be specified directly. The commands for this purpose are based on the FM function.
The command format is:
Syntax:
[AE or BE] FM [v] EN
[AE or BE] prefix specifies Sensor A or Sensor B
FM specifies manual averaging
[v] has allowable values of 0 through 10. Each value represents a particular averaging number.
The numbers are shown in
EN is a required terminating suffix
Example:
AE FM 2 EN
! SETS AVERAGING NUMBER TO 4
Example 2:
AE FM 8 EN
! SETS AVERAGING NUMBER TO 256
Table 3-16: Number Averaging
Value of
v
Averaging Number
0
1
1
2
2
4
3
8
4
16
5
32
6
64
7
128
8
256
9
512
10
1024
Summary of Contents for 8650B Series
Page 3: ...Series 8650B Universal Power Meters Publication 31470 001 Rev C November 2 2017 iii ...
Page 18: ......
Page 32: ......
Page 78: ......
Page 236: ......
Page 244: ......
Page 258: ......
Page 267: ...Publication 31470 001 Rev C November 2 2017 Last page of the document ...