Front Panel Operation
Publication 31470-001, Rev C, November 2, 2017
2-11
b.
External Trigger
Figure 2-6 illustrates the Time Gated measurement parameters with an external trigger. When an
external trigger is input (Point A in Figure 2-6), it starts the Trigger Delay. At the end of the Trigger
Delay, the Gated Time measurement starts and lasts until its preselected time expires. The Holdoff
Time then prevents any further trigger inputs (such as Point B below) from starting a new gated
measurement until it has timed out.
Figure 2-6: External Triggering Illustration