Series 8650B Universal Power Meters
3-90
Publication 31470-001, Rev C, November 2, 2017
The Crest Factor feature returns the current ratio between held power and average power, as displayed on the
front panel. A CRF command does not initiate data collection in same manner as a trigger command, such as
TR1. To get a good reading of the Peak Hold value, the procedure is:
1.
Set up the signal being measured, and send CR1 to reset the Crest Factor measurement.
2.
Send TR2.
3.
Read the TR2 data, or wait for the data ready service request (this allows for settling).
4.
Send CRF.
5.
Read the Crest Factor value.