Remote Operation
Publication 31470-001, Rev C, November 2, 2017
3-143
3.19.17.2
Configuring V
PROP
F
The command format, for purposes of configuring the V
PROP
F feature, is as follows:
Syntax:
VPROPF [A or B] MODE [f] [s]
VPROPF is the function code for the V
PROP
F function
[A or B] specifies Sensor A and Sensor B
MODE indicates that the V
PROP
F feature is being configured (that is, the start frequency and
slope are being specified)
[f] indicates the start frequency (the frequency at zero volts), expressed in Hz. The start
frequency must be less than the upper frequency limit of the sensor
[s] indicates the slope of the V
PROP
F (the ratio of input voltage to input frequency), expressed in
V/Hz. The value must be between 1E-12 and 1E-8
Example:
VPROPF A MODE 2.00E9 1.00E-9
! CONFIGURES THE VPROPF FEATURE FOR SENSOR A AS FOLLOWS:
! 2.00 GHZ START FREQUENCY
! 1.00 VOLT PER GHZ SLOPE