Front Panel Operation
Publication 31470-001, Rev C, November 2, 2017
2-5
2.3.1
Meter Setup
The Meter Setup menu provides the means to configure the meter operating mode, and to store and recall
setups. From the Main menu, press
[Meter Setup]
to display the Meter Setup menu. Press the softkey for the
function to perform. The options are:
•
Calibrator to configure and to turn the Calibrator Output ON or OFF
•
Config to configure the operating mode and parameters
•
Sto/Rcl to store and recall meter setups
•
Service to clear all memory (Additional service routines may be added to future revisions of this
product)
•
Display Return to return to the Main menu without selecting a configuration option
2.3.1.1
Calibrator
The section describes only how to use the Calibrator output as an RF output; See Section 2.4.3 for procedures to
zero and calibrate power sensors. The Calibrator function must be turned ON to use the output as an RF source,
indicated by the words
CAL ON
in the upper left corner of the Main Menu (See Figure 2-2).
Turn on the 8650B for 15 minutes then cycle the power OFF and ON, this establishes the 1 mW reference for the
instrument firmware.
To turn the Calibrator ON or OFF, press
[Meter Setup]
from the Main menu. This will display the Meter Setup
menu. From the Meter Setup menu, press
[Calibrator]
to display the Calibrator menu. Toggle the Calibrator ON
or OFF by pressing
[On/Off]
.
NOTE:
The wording at the top of the screen will not change while this menu is displayed, but it will be
changed to reflect your selection when returning to the Main Menu.
1.
Press
[Level dBm]
to set the RF output level in dBm. Enter the level with the cursor keys.
2.
Press
[Level mW]
to set the RF output level in power. Enter the level with the cursor keys.
3.
Press
[Frequency]
to set the output reference frequency. Enter the frequency with the cursor keys.
Frequencies other than 50 MHz are available only with Option 12 (See Appendix C).
4.
Press
[OK]
to accept the changes or
[Cancel]
to cancel the changes, and return to the Meter Setup
menu.
Summary of Contents for 8650B Series
Page 3: ...Series 8650B Universal Power Meters Publication 31470 001 Rev C November 2 2017 iii ...
Page 18: ......
Page 32: ......
Page 78: ......
Page 236: ......
Page 244: ......
Page 258: ......
Page 267: ...Publication 31470 001 Rev C November 2 2017 Last page of the document ...