Remote Operation
Publication 31470-001, Rev C, November 2, 2017
3-61
Table 3-7: Reset & Power On Default Commands (w/ CW Power Sensor Initialized)
Command
Default
Minimum
Maximum
MEASure<1 | 2>[:SCALar:POWer]
(current power read)
MEMory[:TABLe]:
CHANnel
1
1
2
SELect
ANALOGout
UNIT
Watts
POWer
1e-11,1e-1
0
1e-1
VOLTage
0,10
0
10
OUTPut:
,
ROSCillator[:STATe]
OFF
[BNC:]ANAlog[:STATe]
OFF
SENSe<1 | 2>:
CONFig
CW
AVERage:
COUNt:
1
1
1024
AUTO
ON
TCONtrol
MOVing
CORRection:
FREQ
5e7
0
4e10
VPROpf[:STATe]
OFF
OFFset:
STATe
OFF
[MAGnitude]
0
-99.999
99.999
TRIGger:
SOURce
CW
LEVel[:MAGnitude]
-20.0
-30.0
20.0
OFFSet[:MAGnitude]
0
-5e-8
104e-3
TOTAl[:MAGnitude]
1e-6
-5e-8
104e-3
DELay:
STATe
ON
[MAGnitude]
1e-6
-5e-8
104e-3
TRIGger:
COUNt
1
0
5000
DELay
0.000
0.000
0.050
MODE
POST
SOURce
IMMediate
POWER ON will override the internal battery back up memory and reset only the following configuration items as
shown below. All other configuration items will remain the same as was true before power OFF.
CALCulate<1 | 2>:MODE
NORMal
,
INITiate:CONTinuous
OFF
,
MEMory[:TABLe]:SELect
ANALOGout
,
TRIGger:
COUNt
1
0
5000
DELay
0.000
0.000
0.050
MODE
POST
SOURce
IMMediate