Remote Operation
Publication 31470-001, Rev C, November 2, 2017
3-125
The Peak Hold feature returns the current held value as displayed on the front panel. A PKH command does not
initiate data collection in same manner as a trigger command, such as TR1. To get a good reading of the Peak
Hold value, the procedure is:
1.
Set up the signal being measured, and send PH1 to reset the Peak Hold measurement.
2.
Send TR2.
3.
Read the TR2 data, or wait for the data ready service request (this allows for settling).
4.
Send PKH.
5.
Read the Peak Hold value.