ACOUSTIC CALIBRATION
Last:2003/01/01 10.00
Calibr.level 94.0dB
Leq
dBC
Turn off the calibra-
tor.
94.0
NEXT
5.
After the acoustic calibration, the
electric calibration
will be automatically started. This
stage of the procedure generates the reference data for the following electrical calibrations.
ACOUSTIC CALIBRATION
Last:2003/01/01 10.00
Calibr.level 94.0dB
Leq
dBC
Confirm calibration?
∆
-0.1
ACOUSTIC CALIBRATION
Last:2003/01/01 10.00
Calibr.level 94.0dB
Waiting for electric
calibration ...
Leq
dBC
∆
12.5
YES
NO
6.
At the end, if the values of the electrical calibration are acceptable, press YES to confirm the
new calibration (LEFT arrow of the keypad); at this point, press NO if you want to cancel
the whole calibration (RIGHT arrow of the keypad).
7.
Finally, the microphone polarization check is carried out. Wait until "Take out the preampli-
fier from the calibrator" appears".