ACOUSTIC CALIBRATION
Last:2003/01/01 10.00
Calibr.level 94.0dB
Take off the
calibrator
ACOUSTIC CALIBRATION
Last:2003/01/01 10.00
Calibr.level 94.0dB
...
Leq
dBC
Checking microphone
polarization
∆
-0.1
EXIT
Leq
dBC
96.1
8.
Take out the preamplifier from the calibrator and press EXIT.
9.
The procedure is over.
If calibration constants were incompatible with a correct working of the instrument, calibration
would fail and the message “Calibration failed! Consult the manual” would appear. In this case, re-
fer to “Troubleshooting” (page 90), and if necessary contact our service department.