![GHM DeltaOHM HD2010UC Operating Manual Download Page 17](http://html.mh-extra.com/html/ghm/deltaohm-hd2010uc/deltaohm-hd2010uc_operating-manual_2220414017.webp)
HD2010UC
- 17 -
V4.2
T
IME
P
ROFILE
M
ODE
(
REQUIRES THE
“A
DVANCED
D
ATA
L
OGGER
”
OPTION
)
This display mode, available if “Advanced Data Logger” is installed, presents the time profile
of the A weighted sound pressure level with FAST time constant (L
AFp
). The integration time is
equal to 1/8s and the last 100 measured samples are displayed.
The HD2010UC sound level meter calculates the sound level 128 times a second and displays the
maximum level at intervals equal to 125ms.
Pressing HOLD, the display update will be stopped; however, the instrument continues meas-
uring and pressing HOLD again will restart display updating.
The HOLD status does not affect neither
Monitor
(continuous printing) nor recording opera-
tions. When the continuous recording is activated with the single integration mode, the integration
time acts like a timer for data acquisition, stopping automatically the measurement when the time is
elapsed.
This screen-page is not recorded in the Auto-Store mode.
D
ISPLAY
D
ESCRIPTION
Fig. 5 - Description of the Profile mode display
The Fig. 5 presents the time profile of the A weighted sound pressure level with FAST time
constant.
The integration interval is shown in the left corner at the bottom of the display. Always at the bot-
tom, in the centre, the display shows the measurement unit and the frequency weighting of the
measuring parameter
The amplitude of the vertical scale of the displayed graph corresponds to 5 divisions. The am-
plitude of each division is called “
scale factor
” of the graph and appears in the middle of the verti-
cal axis. Using the RIGHT (zoom +) and LEFT (zoom -) keys, this parameter is selectable in real
time among 20dB, 10dB or 5dB by division.
Use the UP and DOWN arrows to set the graph full scale with steps equal to the selected scale
factor, starting from the instrument full scale
. In this way, the graph can be shifted UP or DOWN,
depending on the key you have pressed.
An “analogue” bar indicator on the display right side provides the non-weighted instantaneous
level of the input sound pressure level, as for the SLM mode bar.
During
Recording
mode, 8 values per second of the L
AFp
level are stored together with 2 val-
ues per second of the SLM screen sound levels. Likewise, when the
Monitor
function is enabled, 4
values are sent to the interface every 0.5s.
1
The instrument full scale is determined by the selection of the input gain by choosing from the menu: MENU >> In-
strument >> Input gain.