46
ACTIVATION OF THE LOCKING FUNCTION
:
1.
To activate the locking function, press the
LOCK
button
(
T
11
) on the hand control and
hold it down
for approx.
3 seconds.
2.
The backlighting of the hand control increases in intensity.
For confirmation
, the LED above the button of the hand
control lights up.
3.
The drives are now locked.
DE
-
ACTIVATION OF THE LOCKING FUNCTION
:
1.
To release, press the
LOCK
button (
T
11
) on the hand
control
twice in quick succession
and
hold it down
for
approx. 3 seconds.
2.
The backlighting of the hand control will now dim slightly.
3.
The drives are now unlocked.
NOTICE!
To prevent unauthorized use of the 601, always activate the
locking function when leaving the treatment room.
If you should forget to do this, the drives lock automatically after 30
minutes.