XDP Specifications
6-1
January 2008
Performance
Specifications
Accuracy:
±0.5% of full scale
Ambient Temperature Range:
–10° to 60°C (14° to 140°F)
Ambient Relative Humidity (maximum):
35% @60°C, 50% @40°C, 65% @20°C, 75% @0°C
Linearity:
±0.5% of full scale
Reliability:
50,000 hours MTBF minimum
Resolution:
±0.1% of full scale
Response Time:
90% of reading in 50 seconds
Stability:
zero drift less than 0.5% of full scale in 24 hours
Temperature Drift:
less than 0.5% of full scale per 100°F
Physical Specifications
Enclosure:
Panel-Mount, Explosion-Proof
(Class I, Division 1, Groups B, C and D)
Fuse:
1.25 A, 250 V, 5x20 mm, IEC Type T
Power Input:
100-240 VAC; 50/60 Hz; 40 W max.
Power Output to Transmitter:
24.0 ± 2.0 VDC; 1 A max.
Note:
The power supply meets CISPR 55022 and CISPR 55014,
Level B EMI requirements and IEC1010-1 safety standards.